No. 1 |
|||
![]() |
TEMPERATURE Type: Mercury in glass or other reference thermometer with Accuracy: ±0.02șC Resolution: ±0.01șC CONDUCTIVITY 1413 us/cm referred to 25șC A solution preparation is based on the procedure in the HANDBOOK OF CHEMISTRY AND PHYSICS 1963 Chemical Rubber Publishing Company. Conductivity of Standard Solutions using KCl. Preparation: Dissolve 0.7459 grams of potassium chloride (KCl) in distilled water made up to one litre of solution. At 25șC this has a conductivity of 1413 us/cm. SALINITY 35ppt Preparation: Obtain clean seawater and standardise using a laboratory salinometer (eg Yeo-Kal Model 601MKIV) whose accuracy is: Accuracy: 0.003ppt Resolution: 0.001ppt
Alternatively, dissolve 32.4356 grams of KCl in 1 kg of solution for a 35 ppt salinity. Conductivity is very dependent upon temperature. This solution of KCl will only give sufficiently accurate results when its temperature is within about 0.1deg C of 15degC. pH Buffers 4.00 and 10.01 prepared using Standard Methods., Ref: Durst, R.A. 1975 Standard Methods; Reference Materials: Standardization of pH Measurements NBS Spec Publ. 260-53, National Bureau of Standards, Washington D.C. pH
Weight of Chemicals needed/1000mls solution @ 25șC
Alternatively, a chemical supply company can provide standard
buffer solutions.
Reference: Innovative Sensors Inc. Technical Bulletin. ORP or REDOX potential can be reported as either the direct potential or alternatively it can be reported as referred to the hydrogen electrode. Prepare pH buffers of 4.0 and 7.0. Add a "pinch" of quinhydrone crystals to each of the buffers until saturated and stir for 30 seconds to obtain the two calibration points. Buffer 7.0 with quinhydrone crystals
gives a probe output of 86mV (or 295.2 mV when referred to the
hydrogen electrode).
DISSOLVED OXYGEN When a new probe is installed it is advisable to do a two point calibration at 0% and 100%. Once this has been done correctly it is only necessary to calibrate the 100% point for field calibrations. The zero point will rarely change. To obtain 0% use either nitrogen gas in a container (i.e. purge oxygen out) or dissolve 26gm of sodium sulphite and add 0.2 gm of cobalt chloride in 500mls of water. Discard solution oafter 30 minutes as it will not keep. To obtain 100% saturated water aerate a sample of fresh water for about 2 hours. A fish tank air pump and aerator stone is ideal. Note: Be sure that the 100% sample is not contaminated with the 0% solution. Wash the sensor before placing it in 100% saturated water. FORMAZIN STANDARD 200 ntu's Preparation for 200 ntu using Standard Methods. Solution 1: 10gm Hexamethylenetetramine C6H12N4 (analysis quality) is dissolved in distilled water. This solution is made up to 100mls with distilled water. Solution 2: 1.0gm Hydrazinium sulphate N2H6SO4 (analysis quality) is dissolved in distilled water. This solution is made up to 100mls with distilled water. Procedure: Mix 5mls of solution 1 with 5mls of solution
2. Allow to stand for 24 hours at 25 ± 3șC, then make up this
solution to 200mls with distilled water. This is 200 ntu. This solution
has a shelf life of 4 weeks if kept in a dark place.
|