![]() |
||||||||||||||||||||||||||||||||||||
|
No. 19 |
||||||||||||||||||||||||||||||||||||
![]() |
References:
These Redox values should only be used when the sensor (Ag/AgCI 3M KCl) is calibrating with a Redox Standard prepared with Quinhydrone. When calibrating with 4.01 buffer mixture, the reading should be between +170 and +185 mV above the reading in the 7.00 buffer mixture, eg. 295.2 mY + l77mV = 472.2 mV @ 250C. If a short span is found-less than a +170 mV change between the 7.00 and 4.01 buffers - the platinum surface may be coated. To remove the coating refer to clause 11.0 “General Care and Maintenance”, pH/ORP Sensor Maintenance. Electrochemical measurements are ultimately referred to the
so-called hydrogen scale (or Normal Hydrogen Electrode NHE), the
convention for which is that the electrochemical potential of a hydrogen
electrode in contact with hydrogen gas at one atmosphere partial pressure
and a solution containing hydrogen ions at unit activity is zero at all
temperatures. The T-611 reference electrode used is the 3M KCl Silver
Chloride type, and this exhibits potential on the hydrogen scale
of:
To refer a
reduction-oxidation potential value measured with a 3M KCL Silver Chloride
electrode to the hydrogen scale, the appropriate value above should be
added to the measured value:
All potentials quoted are
referred to the standard (“normal”) hydrogen electrode (NHE) and are
expressed in mV (reference: International Standard IEC 60746-5,
Oxidation-Reduction Potential or Redox Potential, Annex B).
This document also available in pdf format
|
|||||||||||||||||||||||||||||||||||