Sensor

Range

Accuracy

Resolution

Type

T-611

T-612

T-613

T-617

Temperature

-2 to 50ºC

±0.05ºC

0.01ºC

pt100 element in a stainless steel sleeve.

x

x

x

x

Conductivity

0 to 80 ms/cm
0 to 8000 us/cm

±0.05 ms/cm
±5 us/cm

0.02 ms/cm
3 us/cm

Four platinum electrode cell. Low range uses KCl algorithms, high range uses standard seawater 1 algorithms

x

x

x

x

Salinity 1

0 to 60 ppt

±0.05 ppt

0.02 ppt

Calculated from conductivity and temperature.

x

x

x

x

Dissolved Oxygen

0 to 200% saturation
0 to 20 mg/l

±0.5%

0.1%

Active Pb/Ag cell with a stabilised PTFE membrane, inbuilt stirrer to allow constant flow over membrane for stable reading.

x

x

x

 

Turbidity

0 to 600 ntu

±0.5 up to 300
±5% from  300

0.2 ntu
0.3 ntu

Nephelometric measurement from a detector aligned 90º to the pulsed infra-red source that is unaffected by daylight.

x

x

x

 

pH

0 to 14

±0.03

0.01

Combination Ag/AgCl with junction (common with ORP).

x

x

x

OPT

ORP 2
(REDOX)

-700 mV  to
+1100 mV

±3 mV

1 mV

Combination bare metal electrode with common reference junction with pH sensor.

x

x

x

 

Depth

0 to 100 m or
0 to 150 m

±0.1%
of full scale

0.1 m

Dual active arm silicon strain gauge. Optional atmospheric pressure compensation.

OPT

x

OPT

x

Storage capacity for programmed logging – number of scans 3

2135

16300

 

3171

Storage capacity for operator STORE button – number of scans 3

600

 

 

890

Data Tag – available for operator to insert when using STORE button

 

 

 

50

OPT - Optional

1          Calculated from conductivity and temperature using the Practical Salinity Scale

2           REDOX output is referenced to hydrogen ion electrode as per IEC746-5.

3          Each scan includes a reading of each sensor plus date/time